4-32 MAC
TM
600 Revision G
2047426-002
Maintenance: Check out Procedure
for normal operation.
10. To verify the correctness of programming, go to “Testing the System
Information” on page 3-6 and check if the displayed ARM Software
Version is the latest or not.
BOOT Software Update
1. Select Setup.
2. Enter set up password 1111.
3. Press Enter.
4. Press the Lead and F2 keys together.
5. Enter the service password 7763.
6. Press Enter.
7. Select Update Primary Boot using the down arrow key.
8. Select Yes using the Enter and arrow keys.
A new window opens, askeing you to confirm whether to continue or
not.
9. Select Yes using the arrow and Enter keys.
10. Select Return using the arrow key and press Enter.
11. The system displays the following messages:
Programming Primary boot
Primary boot programming Over
12. Press the esc key to exit service only Setup.
NOTE
Boot SW update is applicable only if the Boot version has changed
from the previous version in the device and is not already
programmed during ARM software update.
System Log
1. Select Setup.
2. Enter the set up password 1111.
3. Press Enter.
4. Press the Lead and F4 keys together.
5. Enter service password 7763.
6. Insert the blank SD card without write protection enabled.
7. Press the Enter key to store the Sys log or the esc key to exit the
screen.
NOTE
The SD card needs to be formatted in FAT 16 file system before use
(manufacturer validated SD card is 2GB Transcend and SanDisk). If