Revision G MAC
TM
600 4-3
2047426-002
Maintenance: Preparing System for FRU Replacement
Preparing System for FRU Replacement
Prior to performing any disassembly procedure, perform these steps.
CAUTION
Take strict precautions against electrostatic discharge
damage while replacing field replaceable units.
Only approved personnel may conduct repairs.
1. Power off the system.
2. Disconnect the unit from the AC wall outlet.
3. Disconnect the power cord from the ACDC adaptor.
4. Disconnect the patient cable from the unit.
5. Remove the battery as described in “PWA Connector Details” on
page 4-4.
6. When replacing FRUs, implement ESD protection if applicable.
7. Defective or empty LI-Ion batteries should be disposed of in
accordance with the applicable legal stipulations or returned to the
factory.
WARNING
OPERATOR SAFETY — Repairing the device when
powered on may cause injury. Necessary Personal
Protective Equipment are recommended while
performing maintenance.
WARNING
To ensure patient safety, use only parts and accessories
manufactured or recommended by GE Healthcare.