PROTEUS XR/a
GE MEDICAL SYSTEMS Operator Manual
REV 11 DIRECTION 2259724-100
7-12
7-6-5 Changing lamps on the multileaf collimator
The lamp of the multileaf collimator may be changed by the user if occasion
demands
ILLUSTRATION 7-12
AUTOMATIC COLLIMATOR LAMP LOCATION
Switch off the system.
Undo both Allen screws on lamp housing.
Remove lamp housing.
Undo the two Allen contact screws on the lamp.
Replace defective lamp.
Do not touch new lamp with your bare fingers.
Screw the two Allen contact screws tight.
Mount lamp housing and fasten it by retightening both screws.
IF THE HALOGEN LAMP OF THE LIGHT LOCALIZER REMAINS
ILLUMINATED FOR A LONGER PERIOD OF TIME, THE HOUSING MAY
HEAT UP.
PLEASE AVOID TOUCHING THE LAMP HOUSING TO PREVENT BURNS.
ALWAYS USE OEM REPLACEMENT LAMPS FOR THE LIGHT LOCALIZER.
HALOGEN LAMPS, WHICH ARE NOT SHORT-CIRCUIT-PROOF, MAY
BREAK AND RESULT IN INJURIES CAUSED BY BROKEN GLASS. LAMP
TYPE DESCRIPTION: LOOK AT THE LABEL AT THE BACKSIDE OF
COLLIMATOR.
Lamp housing
Two Allen screws
WARNING
WARNING
FOR TRAINING PURPOSES ONLY!
NOTE: Once downloaded, this document is UNCONTROLLED, and therefore may not be the latest revision. Always confirm revision status against a validated source (ie CDL).

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Proteus XR/a and is the answer not in the manual?

GE Proteus XR/a Specifications

General IconGeneral
BrandGE
ModelProteus XR/a
CategoryMedical Equipment
LanguageEnglish

Related product manuals