Care and Maintenance
10-6 Venue 50 – Basic Service Manual
5447566-100 English Rev. 7
How often should maintenance tasks be performed? (continued)
NOTE: If conditions exist which exceed typical usage and patient load,
then it is recommended to increase the periodic maintenance
frequencies.
NOTE: The maintenance may require specialized equipment to
complete.
NOTE: The periodic maintenances are not mandatory. The table above
is for reference only.
Table 10-1: Customer Care Schedule
Service at Indicated
Time Daily Weekly Monthly
Per
Facilities
QA
Program
Notes
Clean Probes •* * or before each use
Inspect AC Mains Cable Mobile Ultrasound
system: Check
Weekly
Inspect Cables and
Connectors
Inspect Battery
Clean Console
Clean LCD
Console Leakage Current
Checks
See Notes Twice Annually
Peripheral Leakage
Current Checks
See Notes Twice Annually
Surface Probe Leakage
Current Checks
See Notes Twice Annually
Measurement Accuracy
Checks
See Notes Twice Annually
Functional Checks See Notes also after corrective
maintenance