System maintenance
Venue 50 – Basic Service Manual 10-13
5447566-100 English Rev. 7
Peripheral/option checks
If any peripherals or options are not part of the system
configuration, the check can be omitted.
See ‘Functional checks’ on page 4-8 for more information.
Mains cable inspection
Cleaning
Table 10-7: Mains Cable Inspection, As Appropriate
Step Item Description
1. Unplug Cord Disconnect the mains cable from the wall and Ultrasound system.
2. Inspect Inspect it and its connectors for damage of any kinds.
3. Verify Verify that the LINE, NEUTRAL and GROUND wires are properly attached
to the terminals, and that no strands may cause a short circuit.
Table 10-8: General Cleaning
Step Item Description
1. Console Remove the battery. Use a fluid detergent in warm water on a soft, damp
cloth to carefully wipe the entire system. Be careful not to get the cloth too
wet so that moisture does not enter the console.
2. Probe Holder Clean probe holders. (they may need to be soaked to remove excess gel).
3. LCD Use a soft, non-abrasive folder cloth or use the clearners listed in Chapter
6 of User Manual. Gently wipe the LCD face. DO NOT use a glass cleaner
that has a hydrocarbon base (such as Benzene, Methy Alcohol or Methy
Ethyl Ketone) on LCD with the filter (anti-glare shield).