Introduction
1-12 Venue 50 – Basic Service Manual
5447566-100 English Rev. 7
“ON” indicates the power on position of the
power switch.
CAUTION: This Power Switch DOES NOT
ISOLATE Mains Supply.
See the Console
Overview section for
location information.
“Protective Earth” indicates the protective
earth (grounding) terminal.
• Inside of AC adapter
with docking station
• docking cart
“Equipotentiality” indicates the terminal to
be used for connecting equipotential
conductors when interconnecting
(grounding) with other equipment.
Connection of additional protective earth
conductors or potential equalization
conductors is not necessary in most cases
and is only recommended for situations
involving multiple equipment in a high-risk
patient environment to provide assurance
that all equipment is at the same potential
and operates within acceptable leakage
current limits. An example of a high-risk
patient would be a special procedure
where the patient has an accessible
conductive path to the heart such as
exposed cardiac pacing leads.
IEC60417-5021
Docking Cart
NRTL Listing and Certification Mark is
used to designate conformance to
nationally recognized product safety
standards. The Mark bears the name and/
or logo of the testing laboratory, product
category, safety standard to which
conformity is assessed and a control
number.
Rating Plate
“Consult accompanying documents“ is
intended to alert the user to refer to the
operator manual or other instructions
when complete information cannot be
provided on the label.
Various
This symbol indicates that waste electrical
and electronic equipment must not be
disposed of as unsorted municipal waste
and must be collected separately. Please
contact an authorized representative of
the manufacturer for information
concerning the decommissioning of your
equipment.
Rating Plate
Table 1-5: Product Icons (Continued)
Label/Icon Purpose/Meaning Location