Electrical power
RequirementParameter
100-120 V AC, 50 or 60 Hz, 4.5 ASupply voltage, mixer
220-240 V AC, 50 or 60 Hz, 3.0 A
Two outlets for the following units:Number of electrical outlets
I/O Cabinet
TCU
TNType of distribution system
Disconnect mixer
The mixer must be turned off and a tag-out device attached onto the I/O Cabinet during
installation. The system power switch must be set in the O OFF position and the power
cord must be disconnected.
Computer
An external computer is not required.
Temperature regulation
For temperature control of the jacketed XDM mixer or XDUO mixer, a temperature control
unit (TCU) is required. See the manufacturer´s manual for specified environmental condi-
tions and electrical power requirements for the TCU.
NOTICE
The temperature control unit must be disconnected before moving
the equipment.
120 Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
5 Installation
5.2 Site requirements