EasyManua.ls Logo

GE Xcellerex XDM Series - Autoclave Probe Sheath Assembly

GE Xcellerex XDM Series
322 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
6.4.3 Autoclave probe sheath assembly
Conditions
The probe sheath assembly with inserted probe must be autoclaved.
The recommended conditions for autoclaving are as follows:
Temperature > 121°C
Time 60 min (minimum time 30 min)
Liquid cycle
NOTICE
The autoclaving temperature must not exceed 130°C.
NOTICE
Do not autoclave ratchet clamps.
Prepare for autoclaving
Before autoclaving the probe sheath assembly, prepare the following:
Disposable pipette
Deionized water
Follow the instructions below to prepare the probe sheath for autoclaving:
ActionStep
Verify that the probe cable is not connected to the probe.1
Draw 1 to 2 mL deionized water into a disposable pipette.2
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD 165
6 Preparation
6.4 Prepare and install the probes
6.4.3 Autoclave probe sheath assembly

Table of Contents

Related product manuals