ActionStep
Use the navigation keys to make selections in the menu:6
a
Select Calibrate and press ENTER (2).
b
Select Sensor 2 and press ENTER.
c
Select Cell K.
d
Use the navigation keys to enter the cell constant value that you
found in the accompanying manual.
e
Press ENTER.
f
Press EXIT (3).
Result: The main screen is displayed. The conductivity probe is ready to be
used.
Rinse the probe with de-ionized water.7
Disconnect the probe cable from the probe.8
This procedure is sufficient for most applications. If additional procedures are needed
(changing an offset or resetting the system), refer to the manufacturer's documentation.
164 Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
6 Preparation
6.4 Prepare and install the probes
6.4.2 Prepare the conductivity probe