Calibrate the pH probe
Perform the autocalibration of the pH probe as follows:
ActionStep
Locate the pH/DO transmitter display on the I/O Cabinet. Press MENU (1).1
1
2
3
Result: A menu is displayed on the transmitter screen.
Use the navigation keys to make selections in the menu:2
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD 159
6 Preparation
6.4 Prepare and install the probes
6.4.1 Calibrate the pH probe