ActionStep
Press MENU (1).4
1
2
3
Result: A menu is displayed on the transmitter screen.
Use the navigation keys to make selections in the menu:5
1
Select Calibrate and press ENTER (2).
2
Select Sensor 1 and press ENTER.
3
Select pH and press ENTER.
4
Select Standardize and press ENTER.
5
Enter the externally measured pH value of your sample and press ENTER.
6
Press EXIT (3) repeatedly until you reach to the main screen.
162 Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD
6 Preparation
6.4 Prepare and install the probes
6.4.1 Calibrate the pH probe