Verify weight monitoring
accuracy
Follow the instructions below.
ActionStep
On the local control software home screen, tap Weight.1
Result: The Weight Detail screen opens.
Tare the mixer weight. See Tare mixer weight, on page 225 for instructions.2
Use water or weights to increase the mixer weight:3
Add a known quantity of water into the disposable bag installed inside
the mixer,
or,
place a known quantity of weights onto the mixer.
Xcellerex XDM 50-1000, XDUO 100-1000 mixers Operating Instructions 29251615 AD 261