Installation 2-5
January 2008
Wiring The Remote
Transmitter - J6
Terminal block J6 contains the analog signal and loop power
connections for the remote transmitter. To wire the transmitter,
complete the following steps:
1. Connect the XDP to the transmitter as follows:
Note: Refer to the transmitter manual for details on wiring the
transmitter’s terminal block(s).
a. Connect pin #1 to the positive terminal of the analog output
from the transmitter.
b. Connect pin #2 to the +24 VDC loop power input on the
transmitter.
c. Connect pin #3 to the return connection of the analog output
from the transmitter.
Proceed to the next section to continue wiring the XDP.
Wiring The XDP Alarm
Relays - J3
Terminal block J3 contains connections for the XDP alarm relays. To
wire these alarms, complete the following steps:
Note: A failsafe alarm is wired to the normally-closed (NC)
contacts, while a non-failsafe alarm is wired to the normally-
open (NO) contacts.
1. Connect the Normal alarm as follows (this alarm is factory-
configured for failsafe operation):
a. Connect pin #1 (NC) to the alarm device input.
b. Connect pin #2 (C) to the alarm device return.
c. Pin #3 (NO) is not used for fail-safe operation
2. Connect the XDP Fault alarm as follows (this alarm is factory-
configured for failsafe operation):
a. Connect pin #4 (NC) to the alarm device input.
b. Connect pin #5 (C) to the alarm device return.
c. Pin #6 (NO) is not used for fail-safe operation