EasyManua.ls Logo

GE XDP-H2 - Performing a Zero Calibration

GE XDP-H2
96 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Setup and Calibration 4-21
January 2008
Performing a Zero
Calibration
To perform a field calibration of the meter, access the main menu of
the User Program (see page 3-4) and proceed as follows:
Note: A “Clear-Cal-Drift” option also appears on the above list.
Note: The H2/CO2 calibration gas is used as an example here.
Cal
Opt Disp If necessary, use the [W] and
[
X] keys to select [Cal] and
press
[ENTER].
Cal
Opt Disp Press [ENTER].
1:Cal
Range
Cal
Use the [W] and [X] keys to
select
[Cal] and press [ENTER].
Slct Cal Menu Item
Range
Cal
If necessary, use the [S] and
[
T] keys to select [Manual Cal]
and press
[ENTER].
Manual Cal
View-Cal-Drift
Gas Curve: Use the [
S] and [T] keys to
select the calibration gas curve
and press
[ENTER].
H2/AIR
H2/CO2
Gas-Cur..Manual-Cal Use the [S] and [T] keys to
select
[Zero] and press the
[ENTER] key.
Zero
Span

Table of Contents