September 2007
7-24 Advanced Programming
TERMINAL Option (cont.) • The Bkgd Input Underrange Error is generated when the
background gas compensation signal falls below the normal range.
• The Bkgd Input Overrange Error is generated when the
background gas compensation signal exceeds the normal range.
• The Pressure Input Underrange Error is generated when the
pressure mV signal falls below -512 mV.
ENABLE/DISABLE ERRORS
Press [N] to skip this error, or
press
[Y] or [Enter] to set it.
Bkgd input underrange?
Bkgd input underrange response:
Press [N] to select the desired
option, then press
[Y] or [Enter] to
confirm your selection and move
to the next prompt.
disable [ENABLE]
ENABLE/DISABLE ERRORS
Press [N] to move on to the next
error.
Bkgd input underrange?
ENABLE/DISABLE ERRORS
Press [N] to skip this error, or
press
[Y] or [Enter] to set it.
Bkgd input overrange?
Bkgd input overrange response:
Press [N] to select the desired
option, then press
[Y] or [Enter] to
confirm your selection and move
to the next prompt.
disable [ENABLE]
ENABLE/DISABLE ERRORS
Press [N] to move on to the next
error.
Bkgd input overrange?
ENABLE/DISABLE ERRORS
Press [N] to skip this error, or
press
[Y] or [Enter] to set it.
Pressure input underrange?
Pressure input underrange response:
Press [N] to select the desired
option, then press
[Y] or [Enter] to
confirm your selection and move
to the next prompt.
disable [ENABLE]
ENABLE/DISABLE ERRORS
Press [N] to move on to the next
error.
Pressure input underrange?