EasyManuals Logo

GE XMO2 User Manual

GE XMO2
196 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #99 background imageLoading...
Page #99 background image
September 2007
6-32 General Programming
DriftCal Enabled and Two
Gases Selected (cont.)
The View Drift Curve option:
This option lets you view the DriftCal zero and span curves that
resulted from the most recent zero gas and span gas calibrations.
If
[Y] was pressed above, the XMO2 displays the current %O2 drift at
the current zero gas oxygen level.
The XMO2 now displays the current %O2 drift at the current span
gas oxygen level.
This completes the prompts for the View Offset option, and the XMO2
returns to the Two-Gas DriftCal menu.
DRIFTCAL MENU
Press [Y] or [Enter] to view the
drift curves, or press
[N] to
proceed to the next menu option.
View Drift Curve?
Drift: x.xx %O2 @ xx.xx %O2
When you are done viewing the
data, press
[Y] or [Enter].
Press [YES]
Drift: x.xx %O2 @ xx.xx %O2
When you are done viewing the
data, press
[Y] or [Enter].
Press [YES]

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE XMO2 and is the answer not in the manual?

GE XMO2 Specifications

General IconGeneral
BrandGE
ModelXMO2
CategoryTransmitter
LanguageEnglish

Related product manuals