September 2007
6-32 General Programming
DriftCal Enabled and Two
Gases Selected (cont.)
The View Drift Curve option:
This option lets you view the DriftCal zero and span curves that
resulted from the most recent zero gas and span gas calibrations.
If
[Y] was pressed above, the XMO2 displays the current %O2 drift at
the current zero gas oxygen level.
The XMO2 now displays the current %O2 drift at the current span
gas oxygen level.
This completes the prompts for the View Offset option, and the XMO2
returns to the Two-Gas DriftCal menu.
DRIFTCAL MENU
Press [Y] or [Enter] to view the
drift curves, or press
[N] to
proceed to the next menu option.
View Drift Curve?
Drift: x.xx %O2 @ xx.xx %O2
When you are done viewing the
data, press
[Y] or [Enter].
Press [YES]
Drift: x.xx %O2 @ xx.xx %O2
When you are done viewing the
data, press
[Y] or [Enter].
Press [YES]