-The two front feet of the sterilizer are raised by 2 to 3 cm, resulting in the front end point being higher than the rear end
point.
-Artificially open the door to the steps:
(1)Open the small cover of the sterilizer door.(Fig.3.1-5)
(2)Insert the hexagon spanner into the door shaft and rotated clockwise until the door is fully opened.(Fig.3.1-6)
(3)Remove the operation panel placed in the chamber and connect the operation panel and the control panel.
The connector is tightened.(Fig.3.1-7)
(4)Fixed operation panel.(Fig.3.1-8)
ATTENTION:
When you meet the power supply to stop,you can also follow the above steps to open the door.Remove the
instruments in the chamber before the power is restored.
However, it is important to confirm that the pressure in the before the above operation can be performed,
chamber is reduced to 0.000bar
Fig.3.1-5 Fig.3.1-6 Fig.3.1-7 Fig.3.1-8