EasyManuals Logo

Getinge FD1800 User Manual

Getinge FD1800
64 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #40 background imageLoading...
Page #40 background image
<Doc_TEC><Doc_503712400><Rel_B><Lang_EN>
Page 40 av 64
Input connections
Line 98, sub-lines 01 - 17 Authorization level: D
This line number is used to tell the software which physical inputs are connected to each
logical input. Some sub-line numbers are not included. They are reserved for future use.
The sub-line numbers correspond to the following logical inputs: The “Physical inputs”
column shows the default connection:
Sub-line No Function Physical input
01 Conn., low level, tank 1
02 Conn. intermediate level tank 2
03 Conn., high level, tank 3
04 Conn., descaler, low level 5
05 Connection descaler empty 16
06 Connection chemical agent low level 5
07 Connection chemical agent empty 16
08 Connection rinse-aid low level 6
09 Connection rinse-aid empty 16
10 Conn., process agent, low level 7
11 Connection process agent empty 16
12 Conn., spray agent, low level 8
13 Connection spray agent empty 16
14 Conn., leakage sensor, waste outlet 4
15 Conn., door closed 11
20 Safety switch door closing (FD1810) 13
21 Pedal (FD1810) 14

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge FD1800 and is the answer not in the manual?

Getinge FD1800 Specifications

General IconGeneral
BrandGetinge
ModelFD1800
CategoryCleaning Equipment
LanguageEnglish

Related product manuals