<Doc_TEC><Doc_503712400><Rel_B><Lang_EN>
Page 40 av 64
Input connections
Line 98, sub-lines 01 - 17 Authorization level: D
This line number is used to tell the software which physical inputs are connected to each
logical input. Some sub-line numbers are not included. They are reserved for future use.
The sub-line numbers correspond to the following logical inputs: The “Physical inputs”
column shows the default connection:
Sub-line No Function Physical input
01 Conn., low level, tank 1
02 Conn. intermediate level tank 2
03 Conn., high level, tank 3
04 Conn., descaler, low level 5
05 Connection descaler empty 16
06 Connection chemical agent low level 5
07 Connection chemical agent empty 16
08 Connection rinse-aid low level 6
09 Connection rinse-aid empty 16
10 Conn., process agent, low level 7
11 Connection process agent empty 16
12 Conn., spray agent, low level 8
13 Connection spray agent empty 16
14 Conn., leakage sensor, waste outlet 4
15 Conn., door closed 11
20 Safety switch door closing (FD1810) 13
21 Pedal (FD1810) 14