PROCESS START
Close and lock the door with the handle. Press the door locking key (7) on
sterilizers with automatic door locking, closing door.
Select the appropriate process with the Program selection key (5).
For Classic display, use the navigator.
Activate Start key (8). For Classic display, use the navigator. A password
(558) is required to run Bowie and Dick, Leaktest and Flash program.
The process starts and the
indicator (2) shows that the process is
running.
When the display indicates “FINISH” or “FINISH SERVICE” and the
indicator (3) lights up, the process is completed, open door.
The
indicator will go off once the door is opened.
!
BETWEEN PROCESSES
Make sure that the items are removed from the sterilizing chamber
immediately after the sterilization process.
Remember when handling goods and loading that the chamber and door
are very warm.
When running a 121 °C program after a 134 °C program, wait about 15
minutes before starting.
Close but do not lock the door.
END OF DAY ROUTINE
Close but do not lock the door. The sterilizer can be left with the electric
power supply on.
Shut off the external water supply.
Do not leave instruments in the chamber for longer periods such as
overnight, which might damage the instrument.
Do not run any processes if the autoclave is left unattended for longer
periods such as overnight.
To be performed according to the quality routines on installation site.
Check that the door is steam tight during the process.
Check that the sterilizer can not be started if the door is not correctly closed
and locked.
SAFETY CONTROL
CAUTION!
If ”FINISH” is not shown in the display, items may have been put in the
chamber by someone else, and have not yet been sterilized.
If the
!
indicator (4) is lighting after the sterilizing process, the load
shall not be considered as sterile and must be re-sterilized.
Items may never be left in the chamber between processes.
Unload the sterilizer only after determining the effectiveness of the cycle.
Unloaded goods can’t be wet after sterilization.
The door and chamber area could be HOT. Wear personal protective
equipment when loading and uloading the chamber.
UNLOADING: