Page 18 of 27
200
150 150 150
3150
(
134 °C)
Pre-pulse
vacuum
Sterilization
time
Drying time
vacuum
200
2000
1000
3000
0
Pressure
Time
Phases
2120
(121 °C)
STANDARD PROCESS
PROCESS DESCRIPTIONS
Description Parameter Range Default
P01 Textiles 134 °C
Pre-pulse vacuum
Pre-pulse positive
Sterilization temperature
Sterilization time
Sterilization pressure
Drying time, vacuum
134-137°C
4-10 min
3-3,3 bar
5-30 min
4
0
134 °C
4 min
3-3,3 bar
10 min
P02 Textiles 121 °C
Pre-pulse vacuum
Pre-pulse positive
Sterilization temperature
Sterilization pressure
Sterilization time
Drying time, vacuum
121-124°C
2-2,3 bar
16-30 min
5-30 min
4
0
121 °C
2-2,3 bar
16 min
10 min
P03 18 minutes 134 °C
Pre-pulse vacuum
Pre-pulse positive
Sterilization temperature
Sterilization pressure
Sterilization time
Drying time, vacuum
134-137°C
3-3,3 bar
18-30 min
5-30 min
4
0
134 °C
3-3,3 bar
18 min
10 min
P04 Bowie & Dick-test
Pre-pulse vacuum
Pre-pulse positive
Sterilization temperature
Sterilization pressure
Sterilization time
Drying time, vacuum
134-137°C
3-3,3 bar
4
0
134 °C
3-3,3 bar
3.5 min
3 min
P06 Leak test Leak rate
0-12,9mbar / 10 min 30 min
P05 Flash
Pre-pulse vacuum
Pre-pulse positive
Sterilization temperature
Sterilization pressure
Sterilization time
134-137°C
3-3,3 bar
4
0
134 °C
3-3,3 bar
3.5 min
P22 Heating-up program
(only double-door)
9 min