The sterilizer is equipped with a set of pre-programmed cycles. The process
times mentioned in all documentation are approximate with factory-set values
and depend on the load. The heavier the load, the longer the process. The
material in the load also affects the process time.
All weight indications of the load include the goods to be sterilized as well as
the weight of racks, trays, containers etc. The lighter the trays etc, the more
goods/weight can be sterilized.
Adjustable parameters
The sterilization and drying times are adjustable, from the minimum (as per
the type test) to higher values depending on cycle type, see following pages.
Adjustments are easily done by the operator from the display and require a
password (558).
P01 Textiles 134 °C
EN 285:2015 type-tested process. For sterilization of small and full: porous
loads, hollow loads, metal and rubber loads, wrapped or unwrapped instru-
ments, goods etc.
Total process time incl. drying
Empty chamber ~22 min
Solid load ~8 kg (average load) ~35 min
Full solid load, 14.3 kg (acc. to type test) ~49 min
Small textile load, 0.25 kg (acc. to type test) ~35 min
P02 Textiles 121 °C
EN 285:2015 type-tested process. For sterilization of small and full: porous
loads, hollow loads, metal and rubber loads, wrapped or unwrapped instru-
ments, goods etc.
Total process time incl. drying
Empty chamber ~35 min
Solid load ~8 kg (average load) ~45 min
Full solid load, 14.3 kg (acc. to type test) ~55 min
Small textile load, 0.25 kg (acc. to type test) ~45 min
P03 18 minutes 134 °C
Wrapped/unwrapped instruments, porous and hollow loads. Please note that
this cycle is a general-purpose cycle, to be confi gured in accordance with lo-
cal requirements and regulations. Getinge assumes no responsibility for the
sterilization results of CJD-related goods.
Total process time incl. drying
Empty chamber ~40 min
Solid load ~8 kg (average load) ~49 min
Full solid load, 14.3 kg ~62 min
Small textile load, 0.25 kg ~53 min