Command ?
Type Immediate
Mode Keypad and Remote
Function Request Mode Status
Response format “K” for keypad, “R” for remote
Default response “K”
Command $
Type Immediate
Mode Keypad and Remote
Function Master reset
Response format $ is echoed
Command I
Type Immediate
Mode Keypad and Remote
Function Request Contact Input Status
Response format
“ab” where
“a” reflects pin 3 (START/STOP input)
“b” reflects pin 1 (CW/CCW input)
“a” and “b” take values of either 0 or 1
“1” is an inactive or high input (open)
“0” is an active or low input (closed)
Default response “11
58