CLEANING & DISINFECTING
In this chapter, the term video cable refers to the cable for the reusable system, and the term Smart Cable
refers to the cable for the single-use system.
IMPORTANT
Because the product will likely be contaminated with human blood or body fluids capable of
transmitting pathogens, all cleaning facilities must be in compliance with (U.S.) OSHA Standard
29 CFR 1910.1030 “Bloodborne Pathogens” or an equivalent standard. For more information, visit
www.osha.gov.
WARNING
Availability of cleaning, disinfection, and sterilization products varies by country, and Verathon
®
is unable to test products in every market. For more information, please contact Verathon
Customer Care at 1.800.331.2313 or your local representative. For contact information, visit
verathon.com/contact-us.
WARNING
This product may only be cleaned, disinfected, or sterilized by using the approved processes
provided in this manual. Cleaning, disinfection, and sterilization methods listed are recommended by
Verathon based on efficacy or compatibility with component materials.
WARNING
The reusable Titanium video laryngoscope is considered a semi-critical device intended to contact
the airway. It must be thoroughly cleaned and undergo high-level disinfection after each use.
WARNING
Ensure that you follow the manufacturer’s instructions for handling or disposing of the cleaning,
disinfection, or sterilization solutions provided in this manual.
WARNING
Do not reuse, reprocess or resterilize single-use components. Reuse, reprocessing or resterilization
may create a risk of contamination of the device, cause patient infection or cross-infection.
WARNING