59
Operations & Maintenance Manual: Product Specifications
ELECTROMAGNETIC COMPATIBILITY
GlideScope Titanium system is designed to be in compliance with IEC 60601-1-2:2007, which contains
electromagnetic compatibility (EMC) requirements for medical electrical equipment. The limits for emissions and
immunity specified in this standard are designed to provide reasonable protection against harmful interference in
a typical medical installation.
The GlideScope Titanium system complies with the applicable essential performance requirements specified in
IEC 60601-1 and IEC 60601-2-18. Results of immunity testing show that the essential performance of the system is
not affected under the test conditions described in the following tables. For more information about the essential
performance of the GlideScope Titanium system, see Essential Performance on page 1.
ELECTROMAGNETIC EMISSIONS
Table 16. Guidance and Manufacturer’s DeclarationElectromagnetic Emissions
The GlideScope Titanium system is intended for use in the electromagnetic environment specified below. The
customer or the user of the GlideScope Titanium system should assure that it is used in such an environment.
EMISSIONS TEST COMPLIANCE ELECTROMAGNETIC ENVIRONMENT – GUIDANCE
RF emissions
CISPR 11
Group 1
The GlideScope Titanium system uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to
cause any interference in nearby electronic equipment.
RF emissions
CISPR 11
Class A
The GlideScope Titanium system is suitable for use in all
establishments other than domestic and those directly connected to
the public low-voltage power supply network that supplies buildings
used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies