RECOMMENDED SEPARATION DISTANCES
Table 18. Recommended Separation Distances between Portable and Mobile RF Communications Equipment
and the GlideScope Titanium System
The GlideScope Titanium system is intended for use in an electromagnetic environment in which radiated
RF disturbances are controlled. The customer or the user of the GlideScope Titanium system can help
prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF
communications equipment (transmitters) and the GlideScope Titanium system as recommended below,
according to the maximum output power of the communications equipment.
RATED MAXIMUM
OUTPUT POWER OF
TRANSMITTER (W)
SEPARATION DISTANCE ACCORDING TO FREQUENCY OF TRANSMITTER (m)
150 kHz to 80 MHz
d=1.2P
80 MHz to 800 MHz
d=1.2P
800 MHz to 2.5 GHz
d=2.3P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in
meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the
maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people.
ACCESSORY CONFORMANCE TO STANDARDS
To maintain electromagnetic interference (EMI) within certified limits, the system must be used with the cables,
components, and accessories specified or supplied by Verathon
®
. For additional information, see the System Parts
& Accessories and Component Specifications sections. The use of accessories or cables other than those specified
or supplied may result in increased emissions or decreased immunity of the system.
Table 19. EMC Standards for Accessories
ACCESSORY MAX LENGTH
AC power cord 4.5 m (15 ft)
DC medical power adapter
HDMI-to-DVI cable 6 m (20 ft)
Video cable (reusable system) 2.2 m (7.2 ft)
Smart Cable (single-use system) 1.6 m (5.2 ft)