47
Operations & Maintenance Manual: Maintenance & Safety
MAINTENANCE & SAFETY
PERIODIC INSPECTIONS
Periodic inspections should be performed to ensure safe and effective operation, in addition to performing
routine inspection by the user before and after every use. It is recommended that an operator familiar with the
instrument perform a full visual inspection of all components at least every three months. The inspector should
check the system for the following:
• External damage to the equipment
• Damage to the power supply or adapter
• Damage to the connectors or cable insulation
Report any suspected defects to Verathon
®
Customer Care at:
Phone: 800.331.2313 (U.S. and Canada)
+1.425.867.1348 (International)
Email: customerservice@verathon.com
ELUTION COMPATIBILITY
For use with GlideScope Titanium reusable blades, Verathon
®
has completed testing of compatibility with a
1%sodium dodecyl sulphate (SDS) solution with pH11.0.
The SDS solution is commonly utilized in Europe as an eluting solution to collect residual protein samples from
medical tools or devices that are cleaned after contacting patient tissue. The protein sample solution is then
examined as a verification of the hospital cleaning process.
The testing concluded that 1% SDS solution with pH 11.0 is chemically compatible with the reusable blades and
gives no adverse results when performing repeated 30-minute soaking for 100cycles.
GLIDESCOPE VIDEO MONITOR BATTERY
Under normal operating conditions, the monitor battery will last 2–3 years, or approximately 500charge and
discharge cycles. For more information about the battery, see Battery Specifications.
The battery is not user-replaceable. In case of battery malfunction, do not attempt to replace the monitor
battery. Any attempts to replace the battery by unauthorized service technicians may cause serious harm to the
user and will void the warranty. Please contact your Verathon Customer Care representative for more information
on battery replacement.