Operations & Maintenance Manual: Cleaning & Disinfecting
PROCEDURE 2. DISINFECT A BLADE, VIDEO CABLE, OR SMART CABLE
Before each use, reusable video laryngoscopes must be high-level disinfected, and video cables and Smart Cables
must be low-level disinfected. Use the following instructions to disinfect the GlideScope Smart Cable, video cable,
or reusable Titanium video laryngoscope. Not all chemical solutions listed in Table 9 are compatible with each
component. Ensure that you select a chemical that is compatible with the component you are disinfecting.
In this procedure, the term pure water refers to water that is suitable for disinfection according to local
regulations and your medical facility.
Do not use metal or abrasive brushes, scrub pads, or rigid tools when disinfecting the reusable video laryngoscope.
The window that protects the camera and light can be scratched, permanently damaging the device.
This product is heat-sensitive, and exposing the components to temperatures in excess of 60°C (140°F) will
cause damage to the electronics.
IMPORTANT
When high-level disinfecting a blade, video cable, or Smart Cable, you may use a Medivators
®
CER
Optima1&2AER, DSD-201AER, or SSD-102AER system, provided you do the following:
Use an approved high-level disinfectant from Table 9.
Use a disinfectant that is compatible with the Medivators
®
system. For more information about chemical
compatibility, contact Medivators.
®
Use the conditions provided in Table 9, such as temperature, exposure, and concentration.
Do not expose the component to temperatures exceeding 60°C (140°F) on any cycle.
Table 9. Disinfection Methods for Video Cables, Smart Cables, and Reusable VideoLaryngoscopes
CHEMICAL
DISINFECTION
LEVEL
COMPONENT CYCLES* CONDITIONS
Clinell
®
Universal
Sanitizing Wipes
Low
Video Cable 1500
Conditioning: N/A
Water temperature: N/A
Exposure: Using fresh towelette(s), wet all surfaces and
allow to remain wet for 6 minutes.
Rinse: N/A. Allow the component to thoroughly air dry.
Smart Cable 1500
Clorox
®
Bleach
Germicidal Wipes
Low
Video Cable 3000
Conditioning: N/A
Water temperature: N/A
Exposure: Using fresh towelette(s), wet all surfaces and
allow to remain wet for 3 minutes.
Rinse: N/A. Allow the component to thoroughly air dry.
Smart Cable 1500
STERIS
®
Revital-Ox
Resert
®
XLHLD
Revital-Ox
Resert
®
HLD/
Chemosterilant
Resert
®
XL HLD
High
Blade 3000
Conditioning: 20°C (68°F) or higher
Water Temperature: 20°C (68°F) or higher
Exposure: Soak for 8 minutes, ensuring that all air bubbles
are removed from the surface of the component.
Rinse: (1) 1-minute immersion with agitation in pure
water. Ensure the connector is properly rinsed.
Video Cable 3000
Smart Cable 1500