38
7. If there is any visible sign of contamination in the connectors, use a long, soft-bristled brush or cotton swab
to remove it.
8. Prepare one of the approved cleaning solutions in Table 8 according to the solution manufacturer’s
instructions.
9. Expose the components to the cleaning solution according to the instructions in Table 8. The exposure
process and times vary depending on the solution and the component.
Note:
If you are using Metrex
®
CaviCide
®
, spray additional solution as needed in order to ensure that the
component remains visibly wet for the duration of the exposure period(s).
If you are using a wipe method, rewipe the component as needed in order to ensure that it remains visibly
wet for the duration of the exposure period. You may use multiple wipes as necessary.
10. If applicable for the cleaning solution, rinse the components according to the instructions in Table 8. The
rinsing process and times vary depending on the solution and the component.
To reduce the risk of cytotoxic residual when cleaning with Metrex
®
CaviCide
®
, thoroughly rinse the
component as instructed in this manual.
WARNING
11. Visually inspect the component for contamination. If there is any sign of contamination, restart the
procedure.
12. Using hospital-grade clean air, which is free from oils and residuals found in common compressed air, blow
out the connectors. This dries the connectors and removes any remaining residuals.
13. Using a clean, lint-free cloth, hospital-grade clean air, or a low-temperature dryer, dry the component.
Note: If you are using a wipe method, allow the component to thoroughly air dry.
14. Examine the component for any signs of damage. Reusable titanium blades should not have any signs of
damage other than minor surface scratches or discoloration of the metal as the result of use. If damage is
present, do not use the component, and contact Verathon
®
Customer Care.
The component should now be clean and free of contamination. Handle the product carefully to avoid
recontamination.
Note: Before each use, reusable video laryngoscopes must be high-level disinfected, and video cables and Smart
Cables must be low-level disinfected.