51
Operations & Maintenance Manual: Product Specifications
PRODUCT SPECIFICATIONS
SYSTEM SPECIFICATIONS, STANDARDS, & APPROVALS
Table 13. Titanium System Specifications
GENERAL SPECIFICATIONS
Classification: Electrical Class II, Applied Part BF
Line voltage: Range: 100240 VAC, 50 and 60 Hz. Connect to a medical-grade power supply
DC power supply: 12V DC, 2.5A max
Fuse: Internal 2.5A hold / 5A trip, 15V max
Ingress protection
against water:
Video monitor IP54
Single-use blade IPX4
GlideScope Smart Cable IPX7
Reusable blade and video cable IPX8
Expected product life:
Single-use blade 1 use or 3-year shelf life
Reusable blade 3 years or 3000 cycles
OPERATING & STORAGE SPECIFICATIONS
Operating Conditions
Temperature: 10 to 35°C (50 to 95°F)
Relative humidity: 0 to 95%
Atmospheric pressure: 540–1060 hPa
Shipping & Storage Conditions
Temperature: -20 to 40°C (-4 to 10F)
Relative humidity: 0 to 95%
Atmospheric pressure: 440–1060 hPa