40
CHEMICAL
DISINFECTION
LEVEL
COMPONENT CYCLES* CONDITIONS
STERIS
®
S40™
or S20™
High
Blade 675
Standard cycles in the following processors:
STERIS
®
SYSTEM 1
®
(outside U.S.)
SYSTEM 1E
®
(in U.S.)
SYSTEM 1 EXPRESS (outside U.S.)
SYSTEM 1 PLUS (outside U.S.)
Video Cable 600
Smart Cable 750
ASP
®
Cidex
®
OPA
High
Blade 3000
Conditioning: 20°C (68°F) or higher
Water Temperature: 20°C (68°F) or higher
Exposure: Soak for 12 minutes, ensuring that all air
bubbles are removed from the surface of the blade.
Rinse: (3) 1-minute immersions with agitation in pure water.
Video Cable 3000
Conditioning: 20°C (68°F) or higher
Water Temperature: 20°C (68°F) or higher
Exposure: Soak for 10 minutes, ensuring that all air
bubbles are removed from the surface of the cable.
Rinse: (3) 1-minute immersions with agitation in pure water.
Smart Cable 1500
Metrex
®
MetriCide
®
OPA
Plus
High
Blade 3000
Conditioning: 20°C (68°F) or higher
Water Temperature: 20°C (68°F) or higher
Exposure: Soak for 12 minutes, ensuring that all air
bubbles are removed from the surface of the blade.
Rinse: (3) 1-minute immersions with agitation in pure water.
Video Cable 3000
Conditioning: 20°C (68°F) or higher
Water Temperature: 20°C (68°F) or higher
Exposure: Soak for 10 minutes, ensuring that all air
bubbles are removed from the surface of the cable.
Rinse: (3) 1-minute immersions with agitation in pure water.
Smart Cable 1500
Metrex
®
MetriCide
®
28
§
High
Blade 3000
Conditioning: 25 ± 2°C (77 ± 4°F)
Water temperature: 3340°C (91–104°F)
Exposure: Soak for 20 minutes, ensuring that all air
bubbles are removed from the surface of the blade.
Rinse: (3) 3-minute immersions in pure water, while
agitating, flushing, and brushing with a sterile,
soft-bristled brush.
Video Cable 3000
Conditioning: 25 ± 2°C (77 ± 4°F)
Water temperature: 3340°C (91–104°F)
Exposure: Soak for 20 minutes, ensuring that all air
bubbles are removed from the surface of the cable.
Rinse: (3) 1-minute immersions in pure water, while
agitating, flushing, and brushing with a sterile,
soft-bristled brush.