Operations & Maintenance Manual: Cleaning & Disinfecting
CHEMICAL
DISINFECTION
LEVEL
COMPONENT CYCLES* CONDITIONS
Sultan
®
Healthcare
Sporox
®
II
§
High Video Cable 3000
Conditioning: 20 ± 2°C (68 ± 4°F)
Water Temperature: 3340°C (91–104°F)
Exposure: Soak for 30 minutes, ensuring that all
air bubbles are removed from the surface of the cable.
Following soak, brush and flush. Repeat 30-minute soak.
Rinse: (3) 3-minute immersions with brushing and
flushing in pure water.
Medivators
®
Rapicide
®
PA
30°C
High
Blade 100
Concentration: 850 ± 100 parts per million
Conditioning: 283C (82–90°F)
Exposure: 5 minutes in a Medivators
®
Advantage Plus AER
reprocessing system with the following configuration:
Hookup: 2-8-002HAN Rev. B
Parameter: 1-24-010 C DISF
Video Cable 100
Smart Cable 100
* Value indicates number of compatibility cycles tested on the component. Exceeding the recommended number of cycles may affect the potential life of
the product.
Clinell
®
Universal Sanitizing Wipes are not available in all geographic regions, such as the United States.
This chemical may cause discoloration of metal components, but the discoloration does not affect system efficacy or functionality.
§ Do not use this solution to disinfect a GlideScope Smart Cable.
1. Ensure the video laryngoscope, video cable, or Smart Cable has been properly cleaned, according to the
procedure Clean a Blade, Video Cable, or Smart Cable.
2. Prepare and condition the disinfection solution according to the solution manufacturer’s instructions and the
conditions stated in Table 9.
3. Disinfect the component according to the conditions stated in Table 9. The exposure process and times vary
depending on the solution and the component.
Note: If you are using a wipe method, rewipe the component as needed in order to ensure that it remains
visibly wet for the duration of the exposure period. You may use multiple wipes as necessary.
4. Rinse the component according to the conditions stated in Table 9. The rinsing process and times vary
depending on the solution and the component.
5. Dry the component by using a sterile cloth, hospital-grade clean air, or a low-temperature dryer.
Note: If you are using a wipe method, allow the component to thoroughly air dry.
6. Examine the component for any signs of damage. Reusable titanium blades should not have any signs of
damage other than minor surface scratches or discoloration of the metal as the result of use. If damage is
present, do not use the component, and contact Verathon
®
Customer Care.
7. Store the component in a clean environment.