44
METHOD LEVEL CYCLES* CONDITIONS
ASP
®
Cidex
®
OPA
High 3000
Conditioning: 20°C (68°F) or higher
Water Temperature: 20°C (68°F) or higher
Exposure: Soak for 12 minutes, ensuring that all air bubbles
are removed from the surface of the component.
Rinse: (3) 1-minute immersions with agitation in pure water.
Metrex
®
MetriCide
®
OPA Plus
High 3000
Conditioning: 20°C (68°F) or higher
Water Temperature: 20°C (68°F) or higher
Exposure: Soak for 12 minutes, ensuring that all air bubbles
are removed from the surface of the component.
Rinse: (3) 1-minute immersions with agitation in pure water.
Medivators
®
Rapicide
®
PA 30°C
High 100
Concentration: 750950 parts per million
Conditioning: 283C (82–90°F)
Exposure: 5 minutes in a Medivators
®
Advantage Plus AER
reprocessing system with the following configuration:
Hookup: 2-8-002HAN Rev. B
Parameter: 1-24-010 C DISF
STERIS
®
S40™
or S20™
High 675
Standard cycles in the following processors:
STERIS
®
SYSTEM 1
®
(outside U.S.)
SYSTEM 1E
®
(in U.S.)
SYSTEM 1 EXPRESS (outside U.S.)
SYSTEM 1 PLUS (outside U.S.)
ASP
®
Hydrogen
Peroxide Gas Plasma
Sterilization 500
STERRAD
®
100S (in U.S.)
STERRAD
®
100S short cycle (outside U.S.)
STERRAD
®
NX standard cycle
STERRAD
®
100NX standard cycle
STERRAD
®
50
STERRAD
®
200 short cycle
STERIS
®
Vaprox
®
HC Sterilization 500
Non-lumen cycle in any STERIS
®
Amsco
®
V-PRO
®
low-temperature sterilization system.
Autoclave
(steam cycle)
Sterilization 300
Temperature: 132–137°C (270279°F)
Exposure: 4–18 minutes
* Value indicates number of compatibility cycles tested. Exceeding the recommended number of cycles may affect the potential life of the
product.
This chemical may discolor metal, but the discoloration does not affect efficacy or functionality.