44
METHOD LEVEL CYCLES* CONDITIONS
ASP
®
Cidex
®
OPA
High 3000
Conditioning: 20°C (68°F) or higher
Water Temperature: 20°C (68°F) or higher
Exposure: Soak for 12 minutes, ensuring that all air bubbles
are removed from the surface of the component.
Rinse: (3) 1-minute immersions with agitation in pure water.
Metrex
®
MetriCide
®
OPA Plus
High 3000
Conditioning: 20°C (68°F) or higher
Water Temperature: 20°C (68°F) or higher
Exposure: Soak for 12 minutes, ensuring that all air bubbles
are removed from the surface of the component.
Rinse: (3) 1-minute immersions with agitation in pure water.
Medivators
®
Rapicide
®
PA 30°C
High 100
Concentration: 750–950 parts per million
Conditioning: 28–32°C (82–90°F)
Exposure: 5 minutes in a Medivators
®
Advantage Plus AER
reprocessing system with the following configuration:
• Hookup: 2-8-002HAN Rev. B
• Parameter: 1-24-010 C DISF
STERIS
®
S40™
or S20™
High 675
Standard cycles in the following processors:
STERIS
®
SYSTEM 1
®
(outside U.S.)
SYSTEM 1E
®
(in U.S.)
SYSTEM 1 EXPRESS (outside U.S.)
SYSTEM 1 PLUS (outside U.S.)
ASP
®
Hydrogen
Peroxide Gas Plasma
Sterilization 500
STERRAD
®
100S (in U.S.)
STERRAD
®
100S short cycle (outside U.S.)
STERRAD
®
NX standard cycle
STERRAD
®
100NX standard cycle
STERRAD
®
50
STERRAD
®
200 short cycle
STERIS
®
Vaprox
®
HC Sterilization 500
Non-lumen cycle in any STERIS
®
Amsco
®
V-PRO
®
low-temperature sterilization system.
Autoclave
(steam cycle)
Sterilization 300
Temperature: 132–137°C (270–279°F)
Exposure: 4–18 minutes
* Value indicates number of compatibility cycles tested. Exceeding the recommended number of cycles may affect the potential life of the
product.
† This chemical may discolor metal, but the discoloration does not affect efficacy or functionality.