9. BOARDS AND ELECTRICAL COMPONENTS
9.4 Downloading Firmware
This section describes how to download firmware (system programs) via the USB interface to the flash
memory provided on the data controller board, to which the boot program has been downloaded.
1. Items required to download firmware
• argos2.X file (Firmware)
Use version 3.0 or later to the new data controller board.
Use version 2.91 or former to the old data controller board.
• CS500/IS200 scanner
• PC (Windows 2000/XP and an operating environment that provides USB 2.0 compatibility)
• OPS112 (Scanning Master 21+ Ver 5.22 or later)
• USB interface cable
2. Procedure
(1) Use the USB cable to connect the CS500/IS200 and PC.
(2) Turn on the power while pressing and holding down the STOP key, the REVERSE key and the
FORWARD key until the green and orange LED come on the power LED.
(3) Turn on the power for your PC and start OPS112.
(4) When OPS112 starts, follow the steps given below to download the firmware.
1. Access the "Help" menu and select "About." At this time, a dialog box will appear.
2. Point to the icon in the window and click while pressing and holding down the "Ctrl" key.
3. Click the "OK" button while pressing and holding down the "Shift" key.
This makes the status available for adjustment.
4. Select "Adjust Scanner" from the "Tools" menu to display the Adjust Scanner window.
5. Select "Update Scanner/Firmware" from the "Scan" menu to display the Download dialog box.
(5) Insert the floppy disk containing the firmware (filename: argos2.X) into the floppy disk drive, click the
"View" button on the Download dialog box, select the floppy disk, and click the "OK" button.
(6) Click the "Start" button to begin downloading.
(7) After the download is complete, a message will appear. At this time, the paper and error LEDs on the
scanner should start flashing simultaneously.
After confirming this, turn off the power.