3-17
Settings and Measurement
<Voltage Ranges>
<Temperature Ranges>
<Humidity Range>
(1)-3 Filter
Sets the filter status. Please set the filter to ON when there is likely to be noise in the input.
Filter operation is on a moving average basis.
Range Maximum SPAN Minimum SPAN Minimum Resolution
(Measurement Range)
20mV -22.000 to +22.000mV 0.200mV 0.001mV
50mV -55.00 to +55.00mV 0.50mV 0.01mV
100mV -110.00 to +110.00mV 1.00mV 0.01mV
200mV -220.00 to +220.00mV 2.00mV 0.01mV
500mV -550.0 to +550.0mV 5.0mV 0.1mV
1V -1.1000 to +1.1000V 0.0100V 0.0001V
2V -2.2000 to +2.2000V 0.0200V 0.0001V
5V -5.500 to +5.500V 0.050V 0.001V
10V -11.000 to +11.000V 0.100V 0.001V
20V -22.000 to +22.000V 0.200V 0.001V
50V -55.00 to +55.00V 0.50V 0.01V
1-5V -5.500 to +5.500V 0.050V 0.001V
Range Maximum SPAN Minimum SPAN (p-p) Measurement Range Minimum
Resolution
K -270 to +2000°C 50°C -200 to +1370°C
J -270 to +2000°C 50°C -200 to +1100°C
T -270 to +2000°C 50°C -200 to +400°C
R -270 to +2000°C 50°C 0 to +1600°C
E -270 to +2000°C 50°C -200 to +800°C
B -270 to +2000°C 50°C +600 to +1820°C 0.1°C
S -270 to +2000°C 50°C 0 to +1760°C
N -270 to +2000°C 50°C 0 to +1300°C
W -270 to +2000°C 50°C 0 to +2000°C
PT100 -270 to +2000°C 50°C -200 to +850°C
JPT100 -270 to +2000°C 50°C -200 to +500°C
PT1000 -270 to +2000°C 50°C -200 to +500°C
Range Maximum SPAN Minimum SPAN (p-p) Minimum Resolution
0 to +110% 1.0% 0.1%
Selection items Description
Off No moving average will be calculated.
2 Calculates the moving average of sampling interval x 2.
5 Calculates the moving average of sampling interval x 5.
10 Calculates the moving average of sampling interval x 10.
20 Calculates the moving average of sampling interval x 20.
40 Calculates the moving average of sampling interval x 40.