3-43
Settings and Measurement
(10)-2 Data Search
This function allows you to perform a search within the captured data using the specified method.
The following methods can be specified.
(10)-3 Statistical Calculation between cursors
Statistical calculation is performed on the replay data between two cursors.
Setting items Description
CH Selects the channel you want to search.
Analog Selects analog channels to be searched.
Pulse Selects pulse channels to be searched.
Logic Selects logic channels to be searched.
Alarm Selects alarm channels to besearched.
<When Analog is selected>
CH Selects the analog channel number to be searched.
Mode Selects the mode used for search.
H
Searches the location where the specified channel's signal rose above the specified level.
L
Searches the location where thespecified channel's signal fell below the specified level.
Level Sets the criteria level to perform a search.
<When Pulse is selected>
CH Selects the pulse channel number to be searched.
Mode Selects the mode used for search.
H
Searches the location where the specified channel's signal rose above the specified level.
L
Searches thelocation where the specified channel's signal fell below the specified level.
Level Sets the criteria level to perform a search.
<When Logic is selected>
CH Selects the logic channel number to be searched.
Mode Selects the mode used for search.
H Searches the location where the specified channel's signal changed from L to H.
L Searches the location where thespecified channel's signal changed from H to L.
<When Alarm is selected>
CH Selects the alarm channel number to be searched.
Mode Selects the mode used for search.
Both Searches thelocation where the specified alarm was generated or cancelled.
H Searches the location where the specified alarm was generated.
L Searches the location where the specified alarm was cancelled.
Setting items Description
Function Off Calculation is not performed.
Average Displays the average value of data between two cursors.
Max Displays the maximum value of data between two cursors.
Min Displays the minimum value of data between two cursors.
Peak Displays the peak value of data between two cursors.
RMS Displays the RMS of data between two cursors.
The following formula is used.
*D:data n:numberofdata
R.M.S
ΣD
2
/