position, or automatically finish the setups above by [AUTO].
6. If the display speed is lower after starting the average sampling mode or setting a
long afterglow time.
It is normal.
7. If waveform display is ladder-like:
(1) This appearance is normal.
(2) The horizontal time base gear may be too low, and the display can be improved
by increasing the horizontal time base to increase the horizontal resolution.
(3) The display type may be “VECTOR”, and the linked line between the sampling
points may cause ladder-like display of the waveform. This problem can be
solved by setup the display type as “POINT” display mode.