28
Operation procedure:
1. Press [REF] to display the menu “REF WAV”.
2. Select the “Signal source” as CH1or CH2.
3. Revolve the vertical [POSITION] and knob [Volt/div] to regulate the vertical position
and the gear of the reference waveform.
4. Select “REF A or “REF B” as storage location of the reference waveform.
5. Press “Save” to save the current screen waveform as a waveform reference.
6. Select “REF A ON” or “REF B ON” to recall the reference waveform, as shown in figure
2-23.
The waveform cannot be saved as reference waveform if it is displayed in an X-Y mode.
The horizontal position and gear of the waveform cannot be regulated at a reference
waveform state.