Operation
Firmware update error
If an error message is shown during the update, make sure that the instrument has power
and that the instrument is connected to the correct port on the PC.
Power failure during update
If a power failure occurred during the update, complete the following procedure.
1. Complete steps 1 to 3 in section 4.4.
2. When the message “Waiting for instrument reply” is shown, remove and connect
the terminal connector on the counter. The instrument power must be applied within
30 seconds of opening the firmware file.
If the update fails again, the instrument must be sent to the factory for repair.