EasyManua.ls Logo

Hach Met One 6000 Series - RS232 Wiring; Pulse Wiring

Hach Met One 6000 Series
70 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
20
Installation
3.6.4 RS232 wiring
Refer to Figure 12 and Table 3 to install a particle counter with RS232 communication.
3.6.5 Pulse wiring
Counters with the pulse output option send a pulse signal when a particle is detected. An
external pulse counter or data acquisition system counts the pulses as particles. Pulse
units have two output channels (Ch 1 and Ch 2). Channel 1 reads the smallest size
particles. Channel 2 can be set for a larger size using the setup utility program. The
status output, when asserted, indicates that the particle counter needs attention.
Pulse units cannot be used in a network configuration. The address must always be set
to 1 (Table 8 on page 33). Refer to Figure 13 and Table 4 to install a particle counter with
pulse output.
Figure 12 Terminal assignments—RS232 communication
Table 3 Terminal assignments—RS232 output
Terminal Assignment
1 (not used)
2 (not used)
3RS232 TX
4 RS232 RX
5NC
6Pump TD
7 Pump RD
8 Common (shield ground)
9 Power source (9–28 VDC, 1 A maximum)
10 Common

Table of Contents

Related product manuals