Hach QbD1200 - User Manual

Hach QbD1200
154 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
DOC026.97.80465
QbD1200
07/2014, Edition 2
User Manual
Manual del usuario
Manuel d'utilisation
Manual do Usuário
取扱説明書
사용 설명서
用户手册
使用手冊

Table of Contents

Questions and Answers

  • T
    Theresa ClineSep 9, 2025
    How to fix slow measurements on Hach QbD1200 Measuring Instruments?
    • J
      Jennifer MoralesSep 9, 2025
      To fix the issue where measurements take too long or the results are consistently too low on your Hach Measuring Instruments, replace the UV reactor, as it may be nearing the end of its life. Also, ensure that you have a sufficient quantity of carrier gas and that all fittings around the reactor are tightened properly.
  • S
    Sarah NealSep 7, 2025
    Why the USB import/export function does not operate with Hach QbD1200?
    • H
      Heidi YoungSep 7, 2025
      If the USB import or export function on your Hach Measuring Instruments isn't working, start by checking if the USB flash drive is properly connected. If it is connected, the drive may be corrupted or not recognized. Try fixing or formatting the USB flash drive, and then reconnect it to the instrument.
  • D
    Donald AndrewsAug 22, 2025
    What to do if USB drive is not working with Hach Measuring Instruments?
    • T
      Tina AndersonAug 22, 2025
      If the USB import or export function isn't working with your USB flash drive on the Hach Measuring Instruments, first, ensure the USB flash drive is properly connected to the instrument. If it is connected, the drive may not be recognized or could be corrupt. Try fixing or formatting the USB flash drive, and then reconnect it to the instrument.
  • A
    amandajohnsonSep 5, 2025
    Why user interface error messages appear during menu changes on Hach Measuring Instruments?
    • R
      Robert FlowersSep 5, 2025
      If user interface error messages are appearing during menu changes on your Hach Measuring Instruments, it indicates an error in the user interface firmware. Turn the instrument power off, and then on again.
  • B
    Brandon StevensSep 4, 2025
    What to do if UV reactor does not work on Hach Measuring Instruments?
    • S
      Susan SnyderSep 4, 2025
      If the UV reactor does not work on your Hach Measuring Instruments, it may be expired or damaged. Replace the UV reactor.
  • K
    Kimberly RobinsonSep 2, 2025
    What to do if the system test fails on Hach Measuring Instruments?
    • K
      Kelly VangSep 3, 2025
      If the system test fails on the Hach Measuring Instruments, it could be due to an error during the system test or because the UV reactor is almost expired. Ensure that the carrier gas supply is correctly connected. Also, replace the UV reactor and make sure there is a sufficient quantity of carrier gas.
  • D
    Douglas WolfSep 1, 2025
    Why significant carryover occurs from sample to sample with Hach Measuring Instruments?
    • J
      Julia RobinsonSep 1, 2025
      If significant carryover occurs from sample to sample on your Hach Measuring Instruments, it means that the reactor is not fully flushed after each measurement. Examine the carrier gas supply.
  • D
    Deanna FrancisAug 31, 2025
    What to do if Hach Measuring Instruments shows that sample is not found?
    • C
      Christopher KennedyAug 31, 2025
      If the sample is not found by the Hach Measuring Instruments, it could be due to several reasons. First, ensure that there is enough solution in the reagent bottle by filling it with solution. Next, check that the sample tube is properly connected to the auto sampler and is correctly placed in the sample flask. Finally, ensure that there is a sample to use for the procedure.
  • J
    Jacqueline FoxAug 30, 2025
    Why measurements on Hach Measuring Instruments show the result as "0"?
    • S
      Stephanie ReedAug 30, 2025
      If measurements frequently show the result as "0" on your Hach Measuring Instruments, it is because the background TOC reading is larger than the sample TOC reading. Fill the reagent bottle with water, then perform a new background qualification procedure.
  • A
    Austin WilkinsonAug 28, 2025
    What to do if Hach Measuring Instruments does not accept new password?
    • R
      Ronald HowardAug 28, 2025
      If the new password is not accepted on your Hach Measuring Instruments, it means the password and the confirmation password do not match. Ensure that you enter the same password in both fields.

Summary

Specifications

General information

Safety information

Use of Hazard Information

Defines DANGER, WARNING, CAUTION, and NOTICE symbols for hazard identification.

Precautionary Labels

Explains common safety symbols found on the instrument and their meanings.

Certification

Details regulatory certifications for the instrument, such as FCC and Canadian standards.

Product overview

Product components

Installation

Installation guidelines

Details environmental requirements and safety precautions for instrument installation.

Plumbing

Add water to the GLS water vessel

Step-by-step guide for adding water to the Gas Liquid Separator water vessel.

Connect to air or carrier gas

Electrical installation

Connect to external devices (optional)

Guidance on connecting optional external devices like keyboards, printers, and monitors.

Connect to an auto sampler

Instructions for connecting the auto sampler to the analyzer.

Connect to power

Steps for connecting the instrument to a power source.

User interface and navigation

Display description

Explanation of the home screen layout and key interface elements.

Startup

Log in as the administrator

Steps to log into the instrument as an administrator.

Operation

Set the system settings

Procedure for configuring system settings such as date, time, language, and display.

Add, change or remove a user

Set the report settings

Configure the measurement settings

Configure the peripheral settings

Measurement

Do a measurement

Steps for initiating a new measurement using the instrument.

Make a sample tray template

Guide to creating custom sample tray configurations for measurements.

Data management

Select a report

How to view, filter, format, and output measurement and system reports.

Do a qualifications procedure

Calibration

Instructions for performing instrument calibration procedures.

SST500

Steps for performing the SST500 qualification test.

Background

Guidance on performing the background qualification procedure.

SDBS

Instructions for performing the SDBS qualification test.

Maintenance

Clean spills

Procedures and safety guidelines for cleaning up spills.

Clean the instrument

Instructions for cleaning the exterior surfaces of the instrument.

Replace an instrument component

Guidance on replacing instrument components, with safety warnings.

Install a firmware update

Prepare for shipment

Troubleshooting

Replacement parts and accessories

Replacement parts

List of approved replacement parts for the QbD1200 with their item numbers.

Accessories

List of available accessories for the QbD1200 with their item numbers.

Hach QbD1200 Specifications

General IconGeneral
DetectorSilicon Photodiode
Operating Temperature0 to 50 °C (32 to 122 °F)
Storage Temperature-20 to 60 °C (-4 to 140 °F)
IP RatingIP67
Measurement PrincipleNephelometry
Light SourceLED
DisplayLCD
Power Supply4 AA batteries or USB power adapter
MethodNephelometric determination of turbidity by measuring the amount of light scattered at a 90° angle to the incident beam. A silicon photodiode detects the light.

Related product manuals