Chapter 7, Troubleshooting and Maintenance 85
Haemonetics
®
TEG
®
6s User Manual P/N 115191-IE, Manual revision: AB
1016 Bubble detected in
channel.
Bubbles have been
detected in one or more
channels.
Remove the cartridge and
rerun the test using a new
cartridge. Verify that the
quantity of blood sample is
sufficient for the test. If the
problem persists, contact
the local Haemonetics
representative.
1030 Pressure leak in
instrument
There was a pressure leak
in the analyzer.
Contact the local
Haemonetics
representative.
1031 Pressure leak in cartridge There was a pressure leak
in the cartridge.
Remove the cartridge and
rerun the test using a new
cartridge. If the problem
persists, contact the local
Haemonetics
representative.
1032 Vacuum leak in instrument There was a vacuum leak
in the analyzer.
Contact the local
Haemonetics
representative.
1033 Vacuum leak in cartridge There was a vacuum leak
in the cartridge.
Remove the cartridge and
rerun the test using a new
cartridge. If the problem
persists, contact the local
Haemonetics
representative.
1034 Regulated vacuum error
in cartridge
The regulated vacuum
was out of range when
the cartridge was
clamped.
Remove the cartridge and
rerun the test using a new
cartridge. If the problem
persists, contact the local
Haemonetics
representative.
1037 Regulated pressure error
in instrument
The regulated vacuum
was out of range when
the cartridge was not
clamped.
Contact the local
Haemonetics
representative.
2001 Front panel LED failure The front panel LED failed. The analyzer will continue
to operate correctly.
Report the failure to the
local Haemonetics
representative.
Error Code LCD Display Message Probable Cause Recommended Action