88 Chapter 7, Troubleshooting and Maintenance
P/N 115191-IE, Manual revision: AB Haemonetics
®
TEG
®
6s User Manual
3032 Instrument program error A firmware file cannot be
located by the system.
Update the system
firmware or reset the
system to the previous
firmware revision.
Note: This action
requires
administrator
access; if no
administrator is
available or if the
problem persists,
contact the local
Haemonetics
representative.
3042 Initial Frequency Error Starting frequency
outside of allowed limits.
Remove the cartridge and
rerun the test using a new
cartridge. If the problem
persists, contact the local
Haemonetics
representative.
3054
Error reading barcode. Barcode cannot be read
by analyzer.
Remove the cartridge and
rerun the test using a new
cartridge. If the problem
persists, contact the local
Haemonetics
representative.
3056 Attempting to run a Patient
test using a QC test
cartridge
The cartridge test type is
QC but you have
attempted to run a patient
test.
Remove the QC cartridge
and insert a patient test
cartridge.
3060 QC cartridge expiration
date has passed
The system has detected
that the QC cartridge
expiration date has
passed.
Remove the expired QC
cartridge.
Error Code LCD Display Message Probable Cause Recommended Action