Chapter 7, Troubleshooting and Maintenance 89
Haemonetics
®
TEG
®
6s User Manual P/N 115191-IE, Manual revision: AB
3061 Cartridge Store Table
update required. See
Administrator for
assistance.
The QC cartridge may be
expired. An update to the
Cartridge Store Table is
required to determine the
expiration date.
Contact the System
Administrator for
assistance in downloading
a new Cartridge Store
Table.
Note: This action
requires
administrator
access; if no
administrator is
available, contact
the local
Haemonetics
representative. For
more information,
see “Cartridge
Library” on
page 22.
3063 Cartridge Store Table
update required. See
Administrator for
assistance.
The Patient test cartridge
may be expired. An update
to the Cartridge Store
Table is required to
determine the expiration
date.
Contact the System
Administrator for
assistance in downloading
a new Cartridge Store
Table.
Note: This action
requires
administrator
access; if no
administrator is
available, contact
the local
Haemonetics
representative. For
more information,
see “Cartridge
Library” on
page 22.
3069 Sample delivery timeout. Delivery of sample to test
area did not complete in
expected time.
Remove the cartridge and
rerun the test using a new
cartridge. Verify that
volume of blood sample is
sufficient for the test. Fill
up to or above the line
marked on the side of the
cartridge. If the problem
persists, contact the local
Haemonetics
representative.
Error Code LCD Display Message Probable Cause Recommended Action