Parameter Description Choices Default Basic mode
EC RANGE If AUTO is used,
the edge™ will
automatically
nd the correct
conductivity
range and
unit. If a xed
value is used,
measurements
made outside
the ranges are
considered
Out of Range,
but units will
remain xed
throughout the
measurement.
AUTO,
29.99 μS/cm,
299.9 μS/cm,
2999 μS/cm,
29.99 mS/cm,
200.0 mS/cm,
500.0 mS/cm
AUTO Not available but
measurement
autoranges as
needed.
TDS RANGE If AUTO is used,
the edge™ will
automatically
nd the
correct TDS
measurement
range and
units. If a xed
value is used,
measurements
made outside
the ranges are
considered
Out of Range,
but units will
remain xed
throughout the
measurement.
AUTO,
14.99 mg/L,
149.9 mg/L,
1499 mg/L,
14.99 g/L,
100.0 g/L,
400 g/L
AUTO Not available but
measurement
autoranges as
needed.
TDS UNIT Select units of
measurement
for TDS
measurement.
mg/L or ppm ppm Available to
select.
EC Meter
Conguration

Related product manuals