BENTOCHECK (PROTEIN STABILITY TEST)
Press READ and “----” will blink on the display. After a few seconds the instrument displays
the first reading. The meter continues updating the readings until the “Lamp and Read Status”
indicator turns off. The final displayed value is the averaged reading.
At the end of the measurement, the instrument directly displays turbidity in NTU. This is T1. Record
the value.
e.g.
T1
Use the 1000 µL automatic pipette to add exactly 1 mL of HI83749-0 Bentocheck reagent to
the cuvette. For a correct use of the automatic pipette please follow the related Instruction Sheet.
• Replace the cap. Invert several times to mix and then wait for 1 minute.
• Insert the cuvette into the instrument and close the lid.