BENTOCHECK (PROTEIN STABILITY TEST)
8. BENTOCHECK (PROTEIN STABILITY TEST)
• Turn the instrument ON by pressing ON/OFF. When the LCD displays “----”, the instrument is ready.
Note: On the secondary LCD the current time appears, if selected in Setup menu.
• Select the AVG mode by pressing the AVGkey. The AVG icon will appear on the display.
• Fill a clean, dry cuvette with 10 mL of wine up to the mark, taking care to handle the cuvette by
the top. Replace the cap and wipe the cuvette thoroughly with a lint-free cloth (see General tips for
an accurate measurement).
• Place the cuvette into the instrument and close the lid.