EasyManua.ls Logo

Hanna Instruments HI 83749 - Page 27

Hanna Instruments HI 83749
52 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
27
27
DETERMINATION OF BENTONITE REQUIREMENT
Remove cuvette #1 and open the cap. Use the 1000 µL automatic pipette to add exactly 1 mL
of HI83749-0 Bentocheck reagent to the cuvette. For a correct use of the automatic pipette please
follow the related Instruction Sheet.
• Replace the cap. Invert several times to mix and wait for 1 minute.
Insert the cuvette into the instrument and close the lid. Press READ and “----” will blink on the
display. After a few seconds the instrument displays the first reading. The meter continues updating
the readings until the “Lamp and Read Status” indicator turns off.
• The final displayed value is the averaged reading of turbidity in NTU. This is T2. Record the value.
e.g. sample #1
T2

Related product manuals