79 General Settings
Comm Protocol (Communication Protocol)
Option: Modbus RTU
Read‑only parameter that indicates supported remote communication mode.
With protocol selected, press Setup to start configuring.
Configurable Comunication Protocol Parameters
Net Address
Option: 01 to 99
This option allows the user to set the controller’s Modbus address.
• With item selected press Set to modify.
• Press the
ud keys to increase or decrease the value (keep the key pressed to fast‑forward).
• Press CFM to save.
Baud Rate
Option: 9600, 19200, 38400, 57600, 115200, 256000
This option allows the user to set the desired speed for the serial communication (baud rate) in bps.
Note: The controller and the Modbus server must have the same baud rate.
• With item selected, press Modify for the drop‑down list to display.
• Use the
ud keys to navigate between options.
• Press Select to save.
Parity
Option: None, Odd, Even
This option allows the user to set communication parity based on the parity mode of the connected device.
• With Parity selected, press Modify for the drop‑down list to display.
• Press the
ud keys to navigate between options.
• Press Select to save.