14. ERROR MESSAGES
The instrument shows clear warning messages (bottom of the screen) when erroneous conditions appear
or when measured values are outside the expected range.
The information below provides an explanation of the errors and warnings, and recommended action to be taken.
Displayed Message Explanation & Recommended Action
Temperature under/over range”
Temperature outside specified range.
Replace probe if necessary.
“Under/over compensation range”
During conductivity calibration, the temperature is under/over the
conductivity calibration solution compensation limit.
”EC under/over range”
Outside range in the corresponding scale.
Ensure the vent holes are completely submerged and the sample is within
specified range.
”Broken electrode”
The meter fails to calibrate or gives faulty readings.
Replace the probe.
”EC out of calibration range” Displayed when the measured value is outside calibration range.
”Factory calibration expired” Contact the Hanna technical support for the periodic factory calibration.
15. ACCESSORIES
CONDUCTIVITY SOLUTIONS
HI7030M or HI7030L 12880 μS/cm standard solution, 250 or 500 mL
HI7031M or HI7031L 1413 μS/cm standard solution, 230 or 500 mL
HI7033M or HI7033L 84 μS/cm standard solution, 230 or 500 mL
HI7034M or HI7034L 80000 μS/cm standard solution, 250 or 500 mL
HI7035M or HI7035L 111800 μS/cm standard solution, 230 or 500 mL
HI7037M or HI7037L 100 % NaCl seawater salinity standard solution, 250 or 500 mL
HI7039M or HI7039L 5000 μS/cm standard solution, 250 or 500 mL
OTHER ACCESSORIES
HI740036P 100 mL beaker (10 pcs.)
HI740037P 20 mL beaker (10 pcs.)
HI764060 Electrode holder
HI900946 115 Vac to24 Vdc power adapter, US plug
HI900947 230 Vac to 24 Vdc power adapter, European plug
HI920016 USB type A to C cable
Error Messages  58

Other manuals for Hanna Instruments HI6321

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI6321 and is the answer not in the manual?

Hanna Instruments HI6321 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI6321
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals