EasyManuals Logo
Home>Hanna Instruments>Measuring Instruments>HI6321

Hanna Instruments HI6321 User Manual

Hanna Instruments HI6321
60 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #8 background imageLoading...
Page #8 background image
5. SPECIFICATIONS
5.1. INSTRUMENT
Conductivity
Range
0.000 to 9.999 μS/cm
10.00 to 99.99 μS/cm
100.0 to 999.9 μS/cm
1.000 to 9.999 mS/cm
10.00 to 99.99 mS/cm
100.0 to 1000.0 mS/cm
Resolution
0.001 μS/cm
0.01 μS/cm
0.1 μS/cm
0.001 mS/cm
0.01 mS/cm
0.1 mS/cm
Accuracy ±1 % of reading or ±0.010 μS/cm, whichever is greater
Cell constant 0.0500 to 200.0000 /cm
Calibration type
Automatic
Manual
Calibration points
Single
Multiple
Calibration solutions
84.00 μS/cm
1.413 mS/cm
5.000 mS/cm
12.88 mS/cm
80.00 mS/cm
111.8 mS/cm
Calibration reminder
Disabled
Daily: 0 min. to 23 hours and 59 min.
Periodic: 1 min. to 30 days, 23 hours and 59 min.
Temperature
compensation
Linear
Natural
Standard
Disabled
Reference
temperature
Range 5.0 to 30.0 °C (41.0 to 86.0 °F)
Resolution 0.1 °C / 0.1 °F
Temperature
coefficient
0.00 to 10.00 %/°C
Info taken from HI5321
unless otherwise stated
Specications  8

Other manuals for Hanna Instruments HI6321

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Hanna Instruments HI6321 and is the answer not in the manual?

Hanna Instruments HI6321 Specifications

General IconGeneral
BrandHanna Instruments
ModelHI6321
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals